Device recall: Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software…
Class I FDA enforcement event 96943. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
Class I, in the words of FDA
a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death
Edwards Lifesciences, LLC started this device recall on 14 May 2025. The firm address in the FDA record is in Irvine, CA. FDA put the recall in Class I. FDA event 96943 has 6 product records. The first FDA enforcement report date is 16 Jul 2025. The record gives the status Ongoing. 81 of the 945 device events that started in 2025 have more product records. The median device event of 2025 has 1 product record. The openFDA file has 945 device events that started in 2025 and 972 that started in 2024.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO,…
Recalling firm
Edwards Lifesciences, LLC, Irvine, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
6 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2080-2025 | Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No. | 27022 | Class I | Ongoing | 16 Jul 2025 |
| Z-2081-2025 | Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No | 13240 | Class I | Ongoing | 16 Jul 2025 |
| Z-2082-2025 | Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No | 6801 | Class I | Ongoing | 16 Jul 2025 |
| Z-2083-2025 | Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No | 847 | Class I | Ongoing | 16 Jul 2025 |
| Z-2084-2025 | Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No | 13240 | Class I | Ongoing | 16 Jul 2025 |
| Z-2085-2025 | Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No | 558 | Class I | Ongoing | 16 Jul 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class I: The highest of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class I
- 13%
- Product records in this event
- 6Median of the firm: 1
- Events of the firm on the register
- 50
Other events of the firm
Official channels
- FDA enforcement report Z-2080-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 96943. https://recallregister.com/fda/i/96943/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO,…
- What quantity of product is in FDA event 96943?
- The event has 6 product records. The first record gives this quantity:
27022
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?