Recall Register

Device recall: The Tri-Flo Subglottic Suction System

Class I FDA enforcement event 80508. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Vyaire Medical started this device recall on 20 Jun 2018. The firm address in the FDA record is in Mettawa, IL. FDA put the recall in Class I. FDA event 80508 has 1 product record. The first FDA enforcement report date is 29 Aug 2018. The record gives the status Terminated, with the termination date 28 Jul 2020. 362 of the 1,074 device events that started in 2018 have more product records. The median device event of 2018 has 1 product record. The openFDA file has 1,074 device events that started in 2018 and 1,091 that started in 2017.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Domestic Distribution: AZ, CA, CT, FL, GA, IA, IL, IN, MA, MD, MN, NJ, OH, OK, SD, TX, WA, WI.

Recalling firm

Vyaire Medical, Mettawa, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 80508: product records
Recall numberProductQuantityClassStatusReport date
Z-2826-2018The Tri-Flo Subglottic Suction System2150 Units TotalClass ITerminated29 Aug 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class I
4%
Product records in this event
1Median of the firm: 2
Events of the firm on the register
12

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 80508. https://recallregister.com/fda/i/80508/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Domestic Distribution: AZ, CA, CT, FL, GA, IA, IL, IN, MA, MD, MN, NJ, OH, OK, SD, TX, WA, WI.
What quantity of product is in FDA event 80508?
The FDA record gives this quantity: 2150 Units Total
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
4% of the 1,074 device events that started in 2018 are in Class I.