Device recall: AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,

Class I FDA enforcement event 93821. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Class I, in the words of FDA

a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death

Vyaire Medical started this device recall on 10 Jan 2024. The firm address in the FDA record is in Mettawa, IL. FDA put the recall in Class I. FDA event 93821 has 9 product records. The first FDA enforcement report date is 6 Mar 2024. The record gives the status Ongoing. 62 of the 972 device events that started in 2024 have more product records. The median device event of 2024 has 1 product record. The openFDA file has 972 device events that started in 2024 and 917 that started in 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa

Recalling firm

Vyaire Medical, Mettawa, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

9 product records in the event.

FDA event 93821: product records
Recall numberProductQuantityClassStatusReport date
Z-1057-2024AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,106,656 unitsClass IOngoing6 Mar 2024
Z-1058-2024AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K800166,078 unitsClass IOngoing6 Mar 2024
Z-1059-2024AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F8,220 unitsClass IOngoing6 Mar 2024
Z-1060-2024AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C23,636 unitsClass IOngoing6 Mar 2024
Z-1061-2024AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K80053,683,407 unitsClass IOngoing6 Mar 2024
Z-1062-2024AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F2,070 unitsClass IOngoing6 Mar 2024
Z-1063-2024AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C24,572 unitsClass IOngoing6 Mar 2024
Z-1064-2024AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K80171,020,330 unitsClass IOngoing6 Mar 2024
Z-1065-2024AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K80041,800,000 unitsClass IOngoing6 Mar 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class I: The highest of the three FDA recall classes.

The event in context

Device events that started in 2024
972
Share of these events in Class I
11%
Product records in this event
9Median of the firm: 2
Events of the firm on the register
12

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 93821. https://recallregister.com/fda/i/93821/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
What quantity of product is in FDA event 93821?
The event has 9 product records. The first record gives this quantity: 106,656 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
11% of the 972 device events that started in 2024 are in Class I.