Recall Register

Food, drugs, devices › Becton Dickinson & Company

Device recall: BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509

FDA enforcement event 79223 · Class I · initiated 22 Mar 2018 · status Terminated · terminated 25 Jul 2022

DeviceVoluntary: Firm initiated4 product recordsFranklin Lakes, NJ

Becton Dickinson & Company (Franklin Lakes, NJ) initiated this device recall on 22 Mar 2018; FDA lists 4 product records under event 79223, classified Class I.

FDA defines Class I as a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. 4% of the 1,074 device enforcement events initiated in 2018 carry the same class.

Status: Terminated, terminated 25 Jul 2022. Initiation: voluntary: firm initiated. Becton Dickinson & Company has 84 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Becton Dickinson & Company, Franklin Lakes, NJ.

Reason for recall

"BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology."

Distribution

"Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1139-2018BD Vacutainer¿ EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI:…1,116,012,050Class ITerminated28 Mar 2018
Z-1140-2018BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)3,956,900Class ITerminated28 Mar 2018
Z-1141-2018BD Vacutainer¿ EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894)64,784,900Class ITerminated28 Mar 2018
Z-1142-2018BD Vacutainer¿ Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI: 50382903678841); 3. 367886 (UDI: 50382903678865); 4. 368040 (UDI: 50382903680400); 5. 366667 (UDI: 50382903666671); 6. 367880 (UDI: 50382903678803)117,816,603Class ITerminated28 Mar 2018

Context

4%of 2018 device events are Class I1,074 events that year
4product records in this eventfirm median: 1
84events on this register for the firm

Other enforcement events for Becton Dickinson & Company

InitiatedProduct (first record)TypeClassRecords
27 Apr 2026BD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8DeviceClass I10
27 Jan 202610mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.DeviceClass II1
14 Oct 2025BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;DeviceClass II1
16 Nov 2023PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number DeviceClass II1
11 Oct 2023BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore TuDeviceClass II3
1 Feb 2023BD Ultra-Fine Insulin Syringe 0.5mL, 6mm, 31G - Insulin Syringes are designed for administering insulin in the treatmentDeviceClass II8
15 Sep 2022BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transportinDeviceClass II1
18 Jul 2022BD PhaSealTM Y-Site Connector (C80)- airtight and leak-proof Closed System Drug Transfer Device (CSTD) that mechanicallyDeviceClass II1
9 May 2022BD Connecta Stopcock with OFF Directed Tap Without Extension Tube (500 eaches/carton) Catalog No. 394910 (US)DeviceClass II11
6 Jan 2022BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a caDeviceClass III1

All 84 events for Becton Dickinson & Company

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology."

What does Class I mean?

FDA classifies recalls by relative degree of health hazard. Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

Where was the product distributed?

FDA's distribution pattern field reads: "Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT,…"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 25 Jul 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.