Recall Register

Food, drugs, devices › Becton Dickinson & Company

Device recall: PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number

FDA enforcement event 93562 · Class II · initiated 16 Nov 2023 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordFranklin Lakes, NJ

Becton Dickinson & Company (Franklin Lakes, NJ) initiated this device recall on 16 Nov 2023; FDA lists 1 product record under event 93562, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Becton Dickinson & Company has 84 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Becton Dickinson & Company, Franklin Lakes, NJ.

Reason for recall

"An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump."

Distribution

"US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0736-2024PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 3065475,085,120 unitsClass IIOngoing24 Jan 2024

Context

88%of 2023 device events are Class II917 events that year
1product record in this eventfirm median: 1
84events on this register for the firm

Other enforcement events for Becton Dickinson & Company

InitiatedProduct (first record)TypeClassRecords
27 Apr 2026BD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8DeviceClass I10
27 Jan 202610mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.DeviceClass II1
14 Oct 2025BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;DeviceClass II1
11 Oct 2023BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore TuDeviceClass II3
1 Feb 2023BD Ultra-Fine Insulin Syringe 0.5mL, 6mm, 31G - Insulin Syringes are designed for administering insulin in the treatmentDeviceClass II8
15 Sep 2022BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transportinDeviceClass II1
18 Jul 2022BD PhaSealTM Y-Site Connector (C80)- airtight and leak-proof Closed System Drug Transfer Device (CSTD) that mechanicallyDeviceClass II1
9 May 2022BD Connecta Stopcock with OFF Directed Tap Without Extension Tube (500 eaches/carton) Catalog No. 394910 (US)DeviceClass II11
6 Jan 2022BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a caDeviceClass III1
27 Sep 2021BD Syringe Tip Cap Bulk Sterile Convenience Pak-1000 units/shelf carton. Used to seal syringes and maintain container clDeviceClass II1

All 84 events for Becton Dickinson & Company

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.