Recall Register

Device recall: GlideScope Go Monitors

Class III FDA enforcement event 88214. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Handheld video monitor failed to meet the labeled IP67 rating which may allow fluid ingress into the monitor resulting in loss of functionality.

Class III, in the words of FDA

a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences

Verathon, Inc. started this device recall on 1 Jul 2021. The firm address in the FDA record is in Bothell, WA. FDA put the recall in Class III. FDA event 88214 has 1 product record. The first FDA enforcement report date is 11 Aug 2021. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, Puerto Rico, SC, SD, TN,…

Recalling firm

Verathon, Inc., Bothell, WA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 88214: product records
Recall numberProductQuantityClassStatusReport date
Z-2172-2021GlideScope Go Monitors8,569Class IIIOngoing11 Aug 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class III: The lowest of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class III
3%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
12

Other events of the firm

Other FDA enforcement events of Verathon, Inc.
InitiatedProduct (first record)TypeClassRecords
29 Apr 2024Brand Name: GlideScope Core 15-inch Monitor Model/Catalog Number: 0570-0404 Product Description: Serialized MonitorDeviceClass II2
29 Sep 2020BladderScan Prime Plus Probe REF 0570-0395 - Product Usage: intended to be used for measuring the urine volume in the blDeviceClass II1
5 Jun 2020GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA:DeviceClass I1
31 Mar 2019GlideScope Video Laryngoscope, Operations & Maintenance Manual, 0900-4940 REV 03DeviceClass II1
3 Mar 2017It is recommended for patients with weight/size between 4-22 lbs (1.8-10 kg). Part Number 0270-0429: GVL 2 Stat (box of DeviceClass II1
29 Jan 2016GlideScope Titanium Single-Use Video Laryngoscope. It is a single-use video laryngoscope system (GlideScope Titanium SysDeviceClass I1
30 Nov 2015BladderScan BVI 9600, Model Numbers: 0270-0451, 0270-0452, 0270-0754, and U270-0451. AortaScan AMI 9700, Model Numbers: DeviceClass II1
30 Nov 2015GlideScope Video Laryngoscope (GVL), Part Number GVL 3 - 0574-0007; GVL 4 - 0574-0001; GVL 5- 0574-0030. The system consDeviceClass II2
13 Apr 2015GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscope iDeviceClass II1
8 May 2013GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage: The GlideSDeviceClass I1

Verathon, Inc.: all 12 events

Official channels

Cite this page

Recall Register. FDA enforcement event 88214. https://recallregister.com/fda/iii/88214/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, Puerto Rico, SC, SD, TN,…
What quantity of product is in FDA event 88214?
The FDA record gives this quantity: 8,569
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
3% of the 900 device events that started in 2021 are in Class III.