Device recall: Gel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat. No. 5104 1 x 2.0 mL…
Class III FDA enforcement event 81303. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Due to a customer experiencing complaints on bone fraction of assay not being in range. The complaint: ~12% lower enzyme activity was noted.
Class III, in the words of FDA
a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences
Helena Laboratories, Inc. started this device recall on 7 Nov 2016. The firm address in the FDA record is in Beaumont, TX. FDA put the recall in Class III. FDA event 81303 has 1 product record. The first FDA enforcement report date is 12 Dec 2018. The record gives the status Terminated, with the termination date 15 Jan 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US (nationwide) to states of: GA, IN, and MI, and countries of: Australia, Brazil, Canada, Japan and Hong Kong.
Recalling firm
Helena Laboratories, Inc., Beaumont, TX.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0529-2019 | Gel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat. No. 5104 1 x 2.0 mL FOR IN-VITRO DIAGNOSTIC USE Gel Alkaline Phosphatase Isoenzyme Control 20mL 001831 Alkaline Phosphatase Isoenzyme Control is to be used as a qualitative and/or quantitative control to aid in the… | 43 kits (6 kits Domestic/37 Internationally) | Class III | Terminated | 12 Dec 2018 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class III: The lowest of the three FDA recall classes.
The event in context
- Device events that started in 2016
- 1,232
- Share of these events in Class III
- 5%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 5
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 4 Apr 2017 | "Free and Total Protein S ELISA Kit" Cat. No. 5292 For use in the quantitation of free and total Protein antigen in citr | Device | Class II | 1 |
| 24 Feb 2017 | Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) | Device | Class II | 1 |
| 14 Oct 2016 | Titan Gel Serum Protein Control: SPE Control Cat. No 5136 1 x 2.0 mL | Device | Class II | 1 |
| 20 Apr 2016 | Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 003366 12/13(2) | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0529-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 81303. https://recallregister.com/fda/iii/81303/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US (nationwide) to states of: GA, IN, and MI, and countries of: Australia, Brazil, Canada, Japan and Hong Kong.
- What quantity of product is in FDA event 81303?
- The FDA record gives this quantity:
43 kits (6 kits Domestic/37 Internationally)
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2016 have the same class?