Recall Register

Device recall: Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The…

Class III FDA enforcement event 80434. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.Sorting Drive software version 4.2 was replaced by 4.2.1 (both versions contain the same option for manual barcode printing). When printing consecutively numbered tube labels using this optional Barcode Counter function, the…

Class III, in the words of FDA

a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences

Beckman Coulter Inc. started this device recall on 24 May 2018. The firm address in the FDA record is in Brea, CA. FDA put the recall in Class III. FDA event 80434 has 7 product records. The first FDA enforcement report date is 7 Nov 2018. The record gives the status Terminated, with the termination date 8 Dec 2022.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland,…

Recalling firm

Beckman Coulter Inc., Brea, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

7 product records in the event.

FDA event 80434: product records
Recall numberProductQuantityClassStatusReport date
Z-0353-2019Intelligent Peri-Analytical Workstation (HW+SW), Catalog #: A89492 (iPAW Software v4.2.1), Beckman Coulter, Inc. The iPAW ((intelligent Peri-Analytical Workstation) can act as: a software client for specimen tracking and workload statistics or a physical workstation for manual sorting and label printing.283Class IIITerminated7 Nov 2018
Z-0354-2019Software Package, Sorting-Drive 4.2 User Kit, Part No. B73488, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products. The system automates the process of sorting, decapping, and archiving samples. The Sorting-Drive 4.2.…25Class IIITerminated7 Nov 2018
Z-0355-2019Software Package, Sorting-Drive 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95347, Manufactured by: Beckman Coulter Biomedical GmbH The Sorting-Drive itself is an optional software to be used with the Automate 2500 Family of products. . The Sorting-Drive 4.2.1 User Kit includes the Sorting-Drive Client…54Class IIITerminated7 Nov 2018
Z-0356-2019Software Package, iPAW 4.2 User Kit, Part No. B73489, Manufactured by: Beckman Coulter Biomedical GmbH The iPAW 4.2 User Kit (PN B73489) includes the iPAW software application CD (version 4.2). This function is primarily used to copy an existing barcode ID to replace the label of an individual sample tube because the…14Class IIITerminated7 Nov 2018
Z-0357-2019Software Package, iPAW 4.2.1 User Kit for Windows XP and Windows 10, Part No. B95343, Manufactured by: Beckman Coulter Biomedical GmbH Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes.29Class IIITerminated7 Nov 2018
Z-0358-2019Sorting-Drive Product Part, Sorting-Drive Plus Kit (Sorting Drive Software v4.2.1), Part No. A89483, Beckman Coulter, Inc. The Sorting-Drive Plus Kit includes the Sorting-Drive 4.2.1 User Kit which includes the Sorting-Drive 4.2.1 Software DVD. . The Sorting-Drive software facilitates the Laboratory Information…495Class IIITerminated7 Nov 2018
Z-0359-2019Sorting-Drive Professional Part, Sorting-Drive Professional Kit, (Sorting Drive Software v4.2.1), Part No. A94144, Beckman Coulter, Inc. The Sorting-Drive Professional Kit (PN A94144) includes the Sorting-Drive 4.2.1 User Kit (PN B95347) which includes the Sorting-Drive 4.2.1 Software DVD (PN B95346). This software…58Class IIITerminated7 Nov 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class III: The lowest of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class III
4%
Product records in this event
7Median of the firm: 1
Events of the firm on the register
216

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 80434. https://recallregister.com/fda/iii/80434/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland,…
What quantity of product is in FDA event 80434?
The event has 7 product records. The first record gives this quantity: 283
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
4% of the 1,074 device events that started in 2018 are in Class III.