Recall Register

Device recall: Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No.…

Class II FDA enforcement event 99861. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Ferno-Washington Inc started this device recall on 28 Apr 2026. The firm address in the FDA record is in Wilmington, OH. FDA put the recall in Class II. FDA event 99861 has 1 product record. The first FDA enforcement report date is 21 Oct 2026. The record gives the status Ongoing. 278 of the 677 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, ID, IL, KY, LA, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of…

Recalling firm

Ferno-Washington Inc, Wilmington, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 99861: product records
Recall numberProductQuantityClassStatusReport date
Z-0145-2027Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821319Class IIOngoing21 Oct 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
677
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
9

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 99861. https://recallregister.com/fda/ii/99861/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
27 Oct 2026
Newest FDA report date on the register
21 Oct 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, ID, IL, KY, LA, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of…
What quantity of product is in FDA event 99861?
The FDA record gives this quantity: 319
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 677 device events that started in 2026 are in Class II.