Device recall: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella…

Class II FDA enforcement event 99747. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Abiomed, Inc. started this device recall on 19 Feb 2019. The firm address in the FDA record is in Danvers, MA. FDA put the recall in Class II. FDA event 99747 has 1 product record. The first FDA enforcement report date is 30 Sep 2026. The record gives the status Ongoing. 360 of the 969 device events that started in 2019 have more product records. The median device event of 2019 has 1 product record. The openFDA file has 969 device events that started in 2019 and 1,074 that started in 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.

Recalling firm

Abiomed, Inc., Danvers, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 99747: product records
Recall numberProductQuantityClassStatusReport date
Z-3251-2026Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella…3,322 unitsClass IIOngoing30 Sep 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
969
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
31

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 99747. https://recallregister.com/fda/ii/99747/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
3 Nov 2026
Newest FDA report date on the register
28 Oct 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.
What quantity of product is in FDA event 99747?
The FDA record gives this quantity: 3,322 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 969 device events that started in 2019 are in Class II.