Recall Register

Device recall: Techno-Path Multichem AE Unassayed Level 1 REF:AE621X

Class II FDA enforcement event 99724. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

TECHNO-PATH MANUFACTURING LTD. started this device recall on 14 Aug 2026. The firm address in the FDA record is in Ballina Tipperary. FDA put the recall in Class II. FDA event 99724 has 3 product records. The first FDA enforcement report date is 28 Oct 2026. The record gives the status Ongoing. 134 of the 697 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution in the state of Massachusetts.

Recalling firm

TECHNO-PATH MANUFACTURING LTD., Ballina Tipperary, Ireland.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 99724: product records
Recall numberProductQuantityClassStatusReport date
Z-0205-2027Techno-Path Multichem AE Unassayed Level 1 REF:AE621X832 unitsClass IIOngoing28 Oct 2026
Z-0206-2027Techno-Path Multichem AE Unassayed Level 2 REF: AE622X123 unitsClass IIOngoing28 Oct 2026
Z-0207-2027Techno-Path Multichem AE Unassayed Level 3, REF:AE623X832 unitsClass IIOngoing28 Oct 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
697
Share of these events in Class II
87%
Product records in this event
3Median of the firm: 3
Events of the firm on the register
3

Other events of the firm

Other FDA enforcement events of TECHNO-PATH MANUFACTURING LTD.
InitiatedProduct (first record)TypeClassRecords
17 Oct 2025Multichem ID-B. Model/Catalog Number: SR103B Intended for use as an unassayed positive qualitative quality control serumDeviceClass II2
25 Aug 2023Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12DeviceClass II3

TECHNO-PATH MANUFACTURING LTD.: all 3 events

Official channels

Cite this page

Recall Register. FDA enforcement event 99724. https://recallregister.com/fda/ii/99724/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
3 Nov 2026
Newest FDA report date on the register
28 Oct 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution in the state of Massachusetts.
What quantity of product is in FDA event 99724?
The event has 3 product records. The first record gives this quantity: 832 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 697 device events that started in 2026 are in Class II.