Device recall: Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 /…
Class II FDA enforcement event 99588. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Beckman Coulter Mishima K.K. started this device recall on 3 Aug 2026. The firm address in the FDA record is in Sunto-Gun. FDA put the recall in Class II. FDA event 99588 has 1 product record. The first FDA enforcement report date is 7 Oct 2026. The record gives the status Ongoing. 278 of the 677 device events that started in 2026 have more product records. The median device event of 2026 has 1 product record.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus,…
Recalling firm
Beckman Coulter Mishima K.K., Sunto-Gun, Japan.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-3332-2026 | Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC)… | 492 units | Class II | Ongoing | 7 Oct 2026 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2026
- 677
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 5
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 30 Jan 2026 | DxC 700 AU, REF: B86444, B86446 | Device | Class II | 2 |
| 30 Apr 2025 | Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 17871 | Device | Class II | 2 |
| 10 Jul 2024 | DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, | Device | Class II | 1 |
| 5 Jun 2024 | DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-3332-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 99588. https://recallregister.com/fda/ii/99588/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 27 Oct 2026
- Newest FDA report date on the register
- 21 Oct 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus,…
- What quantity of product is in FDA event 99588?
- The FDA record gives this quantity:
492 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2026 have the same class?
- 87% of the 677 device events that started in 2026 are in Class II.