Device recall: ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum…
Class II FDA enforcement event 99210. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Tosoh Bioscience Inc started this device recall on 12 Jun 2026. The firm address in the FDA record is in Grove City, OH. FDA put the recall in Class II. FDA event 99210 has 1 product record. The first FDA enforcement report date is 22 Jul 2026. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Global Distribution. US Nationwide.
Recalling firm
Tosoh Bioscience Inc, Grove City, OH.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2746-2026 | ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum on specific Tosoh AIA System analyzers, Model Number 0025204 | 310 units | Class II | Ongoing | 22 Jul 2026 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2026
- 572
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 27
Other events of the firm
Official channels
- FDA enforcement report Z-2746-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 99210. https://recallregister.com/fda/ii/99210/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Global Distribution. US Nationwide.
- What quantity of product is in FDA event 99210?
- The FDA record gives this quantity:
310 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2026 have the same class?