Device recall: Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC…
Class II FDA enforcement event 98197. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Medline Industries, LP started this device recall on 16 Dec 2025. The firm address in the FDA record is in Northfield, IL. FDA put the recall in Class II. FDA event 98197 has 6 product records. The first FDA enforcement report date is 21 Jan 2026. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution.
Recalling firm
Medline Industries, LP, Northfield, IL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
6 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1085-2026 | Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359 | 96 units | Class II | Ongoing | 21 Jan 2026 |
| Z-1086-2026 | Medline medical procedure convenience kits labeled as: 1) LAPAROSCOPY CDS-LF, REF CDS860235L; 2) GENERAL LAPAROSCOPIC, REF CDS930223M; 3) LAP CHOLE CDS, REF CDS983678K; 4) LAPAROSCOPY PELV CDS, REF CDS983679K; 5) LAP CHOLE CDS-LF, REF CDS983752M; 6) LAP CHOLE CDS-LF, REF CDS984016B; 7) LAP CHOLE KIT, REF CDS984020P;… | 41432 units | Class II | Ongoing | 21 Jan 2026 |
| Z-1087-2026 | Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC… | 4536 units | Class II | Ongoing | 21 Jan 2026 |
| Z-1088-2026 | Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU… | 12584 units | Class II | Ongoing | 21 Jan 2026 |
| Z-1089-2026 | Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8)… | 1457 units | Class II | Ongoing | 21 Jan 2026 |
| Z-1090-2026 | Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G | 12 units | Class II | Ongoing | 21 Jan 2026 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class II
- 85%
- Product records in this event
- 6Median of the firm: 5
- Events of the firm on the register
- 66
Other events of the firm
Official channels
- FDA enforcement report Z-1085-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 98197. https://recallregister.com/fda/ii/98197/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution.
- What quantity of product is in FDA event 98197?
- The event has 6 product records. The first record gives this quantity:
96 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 85% of the 945 device events that started in 2025 are in Class II.