Recall Register

Device recall: Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Class II FDA enforcement event 98159. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Maquet Cardiopulmonary Gmbh started this device recall on 9 Jan 2026. The firm address in the FDA record is in Rastatt. FDA put the recall in Class II. FDA event 98159 has 1 product record. The first FDA enforcement report date is 11 Feb 2026. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana,…

Recalling firm

Maquet Cardiopulmonary Gmbh, Rastatt, N/A, Germany.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 98159: product records
Recall numberProductQuantityClassStatusReport date
Z-1300-2026Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;3050 unitsClass IIOngoing11 Feb 2026

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2026
572
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
3

Other events of the firm

Other FDA enforcement events of Maquet Cardiopulmonary Gmbh
InitiatedProduct (first record)TypeClassRecords
31 Jul 2026Percutaneous Insertion Kit (PIK), Guidewire 100 cm, Model Numbers: 701053068 (USA), 701047384 (OUS);DeviceClass II3
27 Jun 2025Sprinter Cart/Sprinter Cart XL Infusion Pole. Component used to hold 2 x 2 kg infusion bags.DeviceClass II1

Maquet Cardiopulmonary Gmbh: all 3 events

Official channels

Cite this page

Recall Register. FDA enforcement event 98159. https://recallregister.com/fda/ii/98159/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana,…
What quantity of product is in FDA event 98159?
The FDA record gives this quantity: 3050 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2026 have the same class?
87% of the 572 device events that started in 2026 are in Class II.