Recall Register

Device recall: Halyard, GENERAL LAPAROSCOPY PACK. Catalog Number: VAST018-10.

Class II FDA enforcement event 98045. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Potential for incomplete seal on header bag.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

AVID Medical, Inc. started this device recall on 14 Nov 2025. The firm address in the FDA record is in Toano, VA. FDA put the recall in Class II. FDA event 98045 has 5 product records. The first FDA enforcement report date is 24 Dec 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US distribution to GA, IL, NE, MD.

Recalling firm

AVID Medical, Inc., Toano, VA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 98045: product records
Recall numberProductQuantityClassStatusReport date
Z-0926-2026Halyard, GENERAL LAPAROSCOPY PACK. Catalog Number: VAST018-10.42 unitsClass IIOngoing24 Dec 2025
Z-0927-2026Halyard, EP LAB PK. Catalog Number: ESJH009-03.240 unitsClass IIOngoing24 Dec 2025
Z-0928-2026Halyard, MIDTOWN GENERAL ARTERIOGRAM. Catalog Number: EUHM009-01.300 unitsClass IIOngoing24 Dec 2025
Z-0929-2026Halyard, MAJOR PACK. Catalog Number: JEMJ10-01.90 unitsClass IIOngoing24 Dec 2025
Z-0930-2026Halyard, HEAD AND NECK TRAY. Catalog Number: EURO016-09.56 unitsClass IIOngoing24 Dec 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class II
85%
Product records in this event
5Median of the firm: 3
Events of the firm on the register
24

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 98045. https://recallregister.com/fda/ii/98045/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US distribution to GA, IL, NE, MD.
What quantity of product is in FDA event 98045?
The event has 5 product records. The first record gives this quantity: 42 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
85% of the 945 device events that started in 2025 are in Class II.