Recall Register

Food, drugs, devices › Bard Peripheral Vascular Inc

Device recall: Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

FDA enforcement event 97864 · Class II · initiated 6 Nov 2025 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordTempe, AZ

FDA enforcement event 97864 is a device recall by Bard Peripheral Vascular Inc of Tempe, AZ, initiated 6 Nov 2025 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Bard Peripheral Vascular Inc has 56 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Bard Peripheral Vascular Inc, Tempe, AZ.

Reason for recall

"Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels."

Distribution

"US Nationwide distribution in the states of CA, MN, IN, MI, LA."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0887-2026Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,1240Class IIOngoing17 Dec 2025

Context

85%of 2025 device events are Class II945 events that year
1product record in this eventfirm median: 1
56events on this register for the firm

Other enforcement events for Bard Peripheral Vascular Inc

InitiatedProduct (first record)TypeClassRecords
12 Dec 2025Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060DeviceClass II1
21 Aug 2025Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the VenclDeviceClass II1
5 Feb 2025Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80DeviceClass I1
30 Sep 2024Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CMDeviceClass II1
4 Sep 2024The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedurDeviceClass II1
2 Aug 2024BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy SysteDeviceClass II1
10 Jul 2024Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual LumenDeviceClass II1
14 Nov 2023BD Recanalization System, REF: BDRECANSYSTEMDeviceClass II1
8 Nov 2023Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14GDeviceClass II1
10 Aug 2023Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016DeviceClass II1

All 56 events for Bard Peripheral Vascular Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution in the states of CA, MN, IN, MI, LA."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.