Recall Register

Food, drugs, devices › Bard Peripheral Vascular Inc

Device recall: Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G

FDA enforcement event 93578 · Class II · initiated 8 Nov 2023 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordTempe, AZ

FDA enforcement event 93578 is a device recall by Bard Peripheral Vascular Inc of Tempe, AZ, initiated 8 Nov 2023 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Bard Peripheral Vascular Inc has 56 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Bard Peripheral Vascular Inc, Tempe, AZ.

Reason for recall

"The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures."

Distribution

"Worldwide - US Nationwide distribution including in the states of PR, IL, MA, GA, KY, CA, TX, PA, IA, MN, NH, MD, FL, ID, IN, AL, VA, CO, WI, NY, NV, OK, OH, WA, MI, MS, AZ, HI, SD, NJ, NE, MT, NC,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-0643-2024Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G…1,243,937Class IIOngoing17 Jan 2024

Context

88%of 2023 device events are Class II917 events that year
1product record in this eventfirm median: 1
56events on this register for the firm

Other enforcement events for Bard Peripheral Vascular Inc

InitiatedProduct (first record)TypeClassRecords
12 Dec 2025Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060DeviceClass II1
6 Nov 2025Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,DeviceClass II1
21 Aug 2025Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the VenclDeviceClass II1
5 Feb 2025Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80DeviceClass I1
30 Sep 2024Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CMDeviceClass II1
4 Sep 2024The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedurDeviceClass II1
2 Aug 2024BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy SysteDeviceClass II1
10 Jul 2024Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual LumenDeviceClass II1
14 Nov 2023BD Recanalization System, REF: BDRECANSYSTEMDeviceClass II1
10 Aug 2023Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016DeviceClass II1

All 56 events for Bard Peripheral Vascular Inc

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide - US Nationwide distribution including in the states of PR, IL, MA, GA, KY, CA, TX, PA, IA, MN, NH, MD, FL, ID, IN, AL, VA, CO, WI, NY, NV, OK, OH, WA, MI, MS, AZ, HI, SD, NJ, NE, MT, NC,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.