Device recall: NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP…
Class II FDA enforcement event 97833. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Stryker Corporation started this device recall on 31 Oct 2025. The firm address in the FDA record is in Portage, MI. FDA put the recall in Class II. FDA event 97833 has 3 product records. The first FDA enforcement report date is 3 Dec 2025. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
Recalling firm
Stryker Corporation, Portage, MI.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0599-2026 | NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP; NN-8003 13x10 MHP; NN-8004 13x13 MHP; NN-8005 15x25 MHP GAAB; NN-8006 11X10 MHP; NN-8007 11x13 MHP; NN-8008 19x28 MHP OI; NN-8010 17x31.5 MHP DECQ; NN-8015 NICO Fluid System (NFS); NN-8017 11x13… | 39,148 | Class II | Ongoing | 3 Dec 2025 |
| Z-0600-2026 | NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber | 39,148 | Class II | Ongoing | 3 Dec 2025 |
| Z-0601-2026 | NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK; | 39,148 | Class II | Ongoing | 3 Dec 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class II
- 85%
- Product records in this event
- 3Median of the firm: 2
- Events of the firm on the register
- 37
Other events of the firm
Official channels
- FDA enforcement report Z-0599-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 97833. https://recallregister.com/fda/ii/97833/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
- What quantity of product is in FDA event 97833?
- The event has 3 product records. The first record gives this quantity:
39,148
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 85% of the 945 device events that started in 2025 are in Class II.