Recall Register

Device recall: Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073…

Class II FDA enforcement event 97714. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Medtronic Sofamor Danek USA Inc started this device recall on 16 Sep 2025. The firm address in the FDA record is in Memphis, TN. FDA put the recall in Class II. FDA event 97714 has 1 product record. The first FDA enforcement report date is 3 Dec 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland,…

Recalling firm

Medtronic Sofamor Danek USA Inc, Memphis, TN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 97714: product records
Recall numberProductQuantityClassStatusReport date
Z-0580-2026Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Number: 6068076. 3. SPACER 6068093 CATALYFT PL SHORT 9MM, Model Number: 6068093. 4. SPACER 6068096 CATALYFT…275,260 unitsClass IIOngoing3 Dec 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class II
85%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
31

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 97714. https://recallregister.com/fda/ii/97714/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and Puerto Rico. The countries of Australia, Austria, Belgium, Canada, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, India, Ireland,…
What quantity of product is in FDA event 97714?
The FDA record gives this quantity: 275,260 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?