Device recall: Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02
Class II FDA enforcement event 97610. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
DJO, LLC started this device recall on 25 Sep 2025. The firm address in the FDA record is in Carlsbad, CA. FDA put the recall in Class II. FDA event 97610 has 1 product record. The first FDA enforcement report date is 3 Dec 2025. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
Recalling firm
DJO, LLC, Carlsbad, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0722-2026 | Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02 | 1,594 | Class II | Ongoing | 3 Dec 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class II
- 85%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 7
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 19 Mar 2026 | Brand Name: DONJOY Product Name: ICEMAN CLASSIC CUBE Model/Catalog Number: 11-0494 Software Version: no Product Descript | Device | Class II | 4 |
| 7 Sep 2021 | GEL CONDUCTOR, Description/REF: 5 LITER W/DISP/4238, 24-8.5OZ BOTTLES/4248, 5 LITER/4266, distributed by DJO, LLC / DJO | Device | Class I | 2 |
| 17 Jan 2019 | Aircast REF 3011 - PL RX only VF-PL Sterile Cuff Pkg., Lot# 418003, Expiration Date: 2020 -12 - 31, Made in Mexico, DJO, | Device | Class II | 1 |
| 25 Jun 2018 | This is an accessory to Intelect Advanced, Intelect Mobile, Intelect Neo, and V-Sonic devices. NYLATEX WRAP - Product Us | Device | Class II | 1 |
| 21 Aug 2015 | Monode, Shortwave Accessory Model #: 02200002 Used in conjunction with the Intelect Shortwave Diathermy 100 and Intelect | Device | Class II | 1 |
| 28 May 2013 | VenaFlow Elite 30B-B and 30BI-B Product Usage: The VenaFlow Elite System is a compression pump used for the prevention o | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0722-2026The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 97610. https://recallregister.com/fda/ii/97610/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
- What quantity of product is in FDA event 97610?
- The FDA record gives this quantity:
1,594
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 85% of the 945 device events that started in 2025 are in Class II.