Food, drugs, devices › Fresenius Kabi USA, LLC
Device recall: Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
FDA enforcement event 97554 is a device recall by Fresenius Kabi USA, LLC of North Andover, MA, initiated 3 Sep 2025 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 85% of the 945 device enforcement events initiated in 2025 carry the same class.
Status: Ongoing. Initiation: voluntary: firm initiated. Fresenius Kabi USA, LLC has 67 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Fresenius Kabi USA, LLC, North Andover, MA.
Reason for recall
"Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation."
Distribution
"US Nationwide distribution in the state of Minnesota."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0378-2026 | Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004. | 6 units | Class II | Ongoing | 5 Nov 2025 |
Context
Other enforcement events for Fresenius Kabi USA, LLC
All 67 events for Fresenius Kabi USA, LLC
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US Nationwide distribution in the state of Minnesota."
Is the recall still ongoing?
FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.