Device recall: OtoLase Starter Kit; UPN: AC-2003748. Indicated for a variety of surgical uses including ablation, coagulation…
Class II FDA enforcement event 97279. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
The potential for unsterilized product within finished product labeled as sterile.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
LUMENIS, LTD. started this device recall on 22 Jul 2025. The firm address in the FDA record is in Yokne'Am Ilit. FDA put the recall in Class II. FDA event 97279 has 1 product record. The first FDA enforcement report date is 27 Aug 2025. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Domestic: PA, WI; International: Germany, India;
Recalling firm
LUMENIS, LTD., Yokne'Am Ilit, N/A, Israel.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2344-2025 | OtoLase Starter Kit; UPN: AC-2003748. Indicated for a variety of surgical uses including ablation, coagulation, incision, excision, and vaporization of soft tissue. | 13 units | Class II | Ongoing | 27 Aug 2025 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2025
- 945
- Share of these events in Class II
- 85%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 3
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 23 Jan 2025 | (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Las | Device | Class II | 1 |
| 17 Nov 2015 | Lumenis M22 System Model Number: GA-0005400 (M22 IPL + YAG Module) with Acne Filter (KT-1014971). | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-2344-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 97279. https://recallregister.com/fda/ii/97279/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Domestic: PA, WI; International: Germany, India;
- What quantity of product is in FDA event 97279?
- The FDA record gives this quantity:
13 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2025 have the same class?
- 85% of the 945 device events that started in 2025 are in Class II.