Recall Register

Device recall: Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an…

Class II FDA enforcement event 97197. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Remote Diagnostic Technologies Ltd. started this device recall on 30 May 2025. The firm address in the FDA record is in Farnborough. FDA put the recall in Class II. FDA event 97197 has 1 product record. The first FDA enforcement report date is 6 Aug 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

Recalling firm

Remote Diagnostic Technologies Ltd., Farnborough, United Kingdom.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 97197: product records
Recall numberProductQuantityClassStatusReport date
Z-2175-2025Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R55Class IIOngoing6 Aug 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class II
85%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
12

Other events of the firm

Other FDA enforcement events of Remote Diagnostic Technologies Ltd.
InitiatedProduct (first record)TypeClassRecords
26 Nov 2025Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBPDeviceClass II1
16 Apr 2025Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (DeviceClass II1
15 Apr 2025Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007DeviceClass II1
28 Oct 2024Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator that can deliver a shock in manual DeviceClass III1
9 May 2024Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator ManualDeviceClass II1
24 Aug 2023Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used WitDeviceClass II1
8 May 2023Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020DeviceClass II1
23 Feb 2023Tempus Pro Patient Monitor is a portable vital signs monitor intended to be used by clinicians and medically qualified pDeviceClass II1
28 Jan 2022Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693DeviceClass II1
25 May 2021Tempus Pro - Patient physiological monitor, arrhythmia detector and alarm. Part No. - 00-1004-R, 00-1007-R, 00-1024-R, 0DeviceClass II1

Remote Diagnostic Technologies Ltd.: all 12 events

Official channels

Cite this page

Recall Register. FDA enforcement event 97197. https://recallregister.com/fda/ii/97197/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.
What quantity of product is in FDA event 97197?
The FDA record gives this quantity: 55
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
85% of the 945 device events that started in 2025 are in Class II.