Device recall: Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)

Class II FDA enforcement event 97075. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

SEDECAL SA started this device recall on 10 Jun 2025. The firm address in the FDA record is in Algete. FDA put the recall in Class II. FDA event 97075 has 46 product records. The first FDA enforcement report date is 29 Oct 2025. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US: CA, IL, and NJ

Recalling firm

SEDECAL SA, Algete, Spain.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

46 product records in the event.

FDA event 97075: product records
Recall numberProductQuantityClassStatusReport date
Z-0242-2026Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)56 unitsClass IIOngoing29 Oct 2025
Z-0243-2026MobileDiagnost wDR 2.2; Model Number REF 9890-010-89522, Mobile X-ray system1198 unitsClass IIOngoing29 Oct 2025
Z-0244-2026Model Number 40KWFX, Mobile X-ray system55 unitsClass IIOngoing29 Oct 2025
Z-0245-2026Model Number 40KW-FX.899, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0246-2026Model Number 40KWFXPLUS, Mobile X-ray system268 unitsClass IIOngoing29 Oct 2025
Z-0247-2026Model Number 40KWFXPLUS.002, Mobile X-ray system8 unitsClass IIOngoing29 Oct 2025
Z-0248-2026Model Number 40KWFXPLUS.004, Mobile X-ray system3 unitsClass IIOngoing29 Oct 2025
Z-0249-2026Model Number 40KWFXPLUS.005, Mobile X-ray system12 unitsClass IIOngoing29 Oct 2025
Z-0250-2026Model Number 40KWFXPLUS.889, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0251-2026Model Number 40KWFXPLUS-710CW, Mobile X-ray system12 unitsClass IIOngoing29 Oct 2025
Z-0252-2026Model Number 40KWFXPLUS-710CW.007, Mobile X-ray system7 unitsClass IIOngoing29 Oct 2025
Z-0253-2026Model Number 40KWMOBILE3.899, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0254-2026Model Number SM-20HF-Batt; 20KW ANALOG, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0255-2026Model Number SM-32HF-B-D-C; 32KW 50G, Mobile X-ray system8 unitsClass IIOngoing29 Oct 2025
Z-0256-2026Model SM-32HF-Batt; 32KW ANALOG, Mobile X-ray system22 unitsClass IIOngoing29 Oct 2025
Z-0257-2026Model SM-40HF-B-D-C; 40KW - CANON READY, Mobile X-ray system6 unitsClass IIOngoing29 Oct 2025
Z-0258-2026Model Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray system39 unitsClass IIOngoing29 Oct 2025
Z-0259-2026Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0260-2026Model SM-40HF-B-D-C; 40KW 70C, Mobile X-ray system14 unitsClass IIOngoing29 Oct 2025
Z-0261-2026Model Number SM-40HF-Batt; 40KW ANALOG, Mobile X-ray system9 unitsClass IIOngoing29 Oct 2025
Z-0262-2026Sedecal Easy Moving Plus Model Number SM-50HF-B-D-C; 50KW 50G, Mobile X-ray system8 unitsClass IIOngoing29 Oct 2025
Z-0263-2026Model Number SM-40HF-B-D-C; 40KW 50C, Mobile X-ray system4 unitsClass IIOngoing29 Oct 2025
Z-0264-2026Model Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray system27 unitsClass IIOngoing29 Oct 2025
Z-0265-2026Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system5 unitsClass IIOngoing29 Oct 2025
Z-0266-2026Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0267-2026Model Number SM-40HF-B-D-C; 40KW 55G, Mobile X-ray system5 unitsClass IIOngoing29 Oct 2025
Z-0268-2026Model Number SM-40HF-B-D-C; 40KW 55G/60G, Mobile X-ray system2 unitsClass IIOngoing29 Oct 2025
Z-0269-2026Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0270-2026Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system4 unitsClass IIOngoing29 Oct 2025
Z-0271-2026Model Number SM-40HF-B-D-C; 40KW 70C, Mobile X-ray system181 unitsClass IIOngoing29 Oct 2025
Z-0272-2026Model Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0273-2026Model Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system10 unitsClass IIOngoing29 Oct 2025
Z-0274-2026Model Number SM-40HF-B-D-C; 40KW 80C, Mobile X-ray system6 unitsClass IIOngoing29 Oct 2025
Z-0275-2026Model Number SM-40HF-B-D-C; 40KW CANON READY, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0276-2026Model Number SM-40HF-B-D-C; 40KW CANON READY.005, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0277-2026Model Number SM-40HF-B-D-C; 40KW, Mobile X-ray system142 unitsClass IIOngoing29 Oct 2025
Z-0278-2026Model Number SM-40HF-B-D-C; 40KW.002, Mobile X-ray system6 unitsClass IIOngoing29 Oct 2025
Z-0279-2026Model Number SM-40HF-B-D-C; 40KW.005, Mobile X-ray system6 unitsClass IIOngoing29 Oct 2025
Z-0280-2026Model Number SM-40HF-B-D-C; 40KW70CDEMO.005, Mobile X-ray system6 unitsClass IIOngoing29 Oct 2025
Z-0281-2026RADPRO MOBILE, Mobile X-ray system24 unitsClass IIOngoing29 Oct 2025
Z-0282-2026SED URS LP, Mobile X-ray systemClass IIOngoing29 Oct 2025
Z-0283-2026SHIMADZU RF.004, Mobile X-ray systemClass IIOngoing29 Oct 2025
Z-0284-2026SM_40HF_B_D_VIR3, Mobile X-ray system17 unitsClass IIOngoing29 Oct 2025
Z-0285-2026SM_40HF_B_D_VIR3.004, Mobile X-ray system3 unitsClass IIOngoing29 Oct 2025
Z-0286-2026SM_40HF_BDC, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025
Z-0287-2026SM_40HF_BDC3, Mobile X-ray system1 unitClass IIOngoing29 Oct 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2025
945
Share of these events in Class II
85%
Product records in this event
46Median of the firm: 1
Events of the firm on the register
7

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 97075. https://recallregister.com/fda/ii/97075/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US: CA, IL, and NJ
What quantity of product is in FDA event 97075?
The event has 46 product records. The first record gives this quantity: 56 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2025 have the same class?
85% of the 945 device events that started in 2025 are in Class II.