Recall Register

Food, drugs, devices › GE Medical Systems, LLC

Device recall: Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile

FDA enforcement event 96027 · Class II · initiated 6 Dec 2024 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordWaukesha, WI

FDA enforcement event 96027 is a device recall by GE Medical Systems, LLC of Waukesha, WI, initiated 6 Dec 2024 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 87% of the 970 device enforcement events initiated in 2024 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. GE Medical Systems, LLC has 42 enforcement events on this register, with a median of 3 product records per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

GE Medical Systems, LLC, Waukesha, WI.

Reason for recall

"There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time."

Distribution

"Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0918-2025Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB7 systemsClass IIOngoing22 Jan 2025

Context

87%of 2024 device events are Class II970 events that year
1product record in this eventfirm median: 3
42events on this register for the firm

Other enforcement events for GE Medical Systems, LLC

InitiatedProduct (first record)TypeClassRecords
5 Jun 2026Optima XR200amx, Mobile Digital-Ready Radiographic SystemDeviceClass II3
1 May 2026Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All ModelsDeviceClass II3
20 Apr 2026Allia IGS 3 Pulse angiographic X-ray systemDeviceClass II5
26 Mar 2026Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300DeviceClass II7
3 Feb 2026Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E8DeviceClass II1
30 Jan 2026Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including DeviceClass II3
24 Oct 2025Senographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and diagnostic mammographyDeviceClass II1
18 Sep 2025LOGIQ P10 series with software version R4.5.7 Model Number 5877534DeviceClass II3
22 Aug 2025MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model NuDeviceClass II1
2 Jul 2025GE HealthCare OEC 9800.DeviceClass II2

All 42 events for GE Medical Systems, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.