Recall Register

Device recall: CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM…

Class II FDA enforcement event 95814. Recall start date: . Status in the FDA record: Completed.

Reason for recall, in the words of FDA

Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Accuray Incorporated started this device recall on 23 Dec 2024. The firm address in the FDA record is in Madison, WI. FDA put the recall in Class II. FDA event 95814 has 1 product record. The first FDA enforcement report date is 5 Feb 2025. The record gives the status Completed.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution.

Recalling firm

Accuray Incorporated, Madison, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 95814: product records
Recall numberProductQuantityClassStatusReport date
Z-1023-2025CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the…212Class IICompleted5 Feb 2025

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
17

Other events of the firm

Other FDA enforcement events of Accuray Incorporated
InitiatedProduct (first record)TypeClassRecords
16 Jun 2026Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear AcceleratorDeviceClass II1
16 Jun 2026Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear AcceleratorDeviceClass II1
3 Aug 2023Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.DeviceClass II1
26 Feb 2021CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precisionDeviceClass II1
10 Dec 2020TomoTherapy Treatment System - Product Usage: used as an integrated system for the planning and precise delivery of radiDeviceClass II2
26 Sep 2018CyberKnife M6, Part Number 054000-004 The CyberKnife M6 Systems are indicated for treatment planning and image guided stDeviceClass II1
31 Aug 2018Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systemsDeviceClass II1
10 Apr 2018CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment DeviceClass II2
17 Nov 2017Hi-Art(R) System, Model Number H-000-0003 Product Usage: The TomoTherapy treatment system is intended to be used as an iDeviceClass II4
6 Dec 2016CyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI. Radiology: The CyberKnife Robotic Radiosurgery System is iDeviceClass II1

Accuray Incorporated: all 17 events

Official channels

Cite this page

Recall Register. FDA enforcement event 95814. https://recallregister.com/fda/ii/95814/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution.
What quantity of product is in FDA event 95814?
The FDA record gives this quantity: 212
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.