Device recall: MICRO-X Rover Mobile X-ray System, # MXU-RV35
Class II FDA enforcement event 95624. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Micro-X Ltd. started this device recall on 22 Oct 2024. The firm address in the FDA record is in Tonsley. FDA put the recall in Class II. FDA event 95624 has 1 product record. The first FDA enforcement report date is 4 Dec 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US: TX, PR, MD, CA
Recalling firm
Micro-X Ltd., Tonsley, Australia.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0580-2025 | MICRO-X Rover Mobile X-ray System, # MXU-RV35 | 6 | Class II | Ongoing | 4 Dec 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 8
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 27 Mar 2026 | Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71 | Device | Class II | 1 |
| 4 Apr 2025 | Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71, MXU-RV35 Product Description: Rover | Device | Class II | 2 |
| 28 Mar 2025 | Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-r | Device | Class II | 2 |
| 23 Oct 2024 | MICRO-X Rover Mobile X-ray System, # MXU-RV19 | Device | Class II | 1 |
| 28 Nov 2023 | MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 | Device | Class II | 1 |
| 31 May 2023 | MICRO-X Rover Mobile X-ray System, # MXU-RV35 | Device | Class II | 1 |
| 31 Aug 2022 | MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0580-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 95624. https://recallregister.com/fda/ii/95624/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US: TX, PR, MD, CA
- What quantity of product is in FDA event 95624?
- The FDA record gives this quantity:
6
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?