Recall Register

Device recall: Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and…

Class II FDA enforcement event 95380. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Olympus Corporation of the Americas started this device recall on 13 Aug 2024. The firm address in the FDA record is in Center Valley, PA. FDA put the recall in Class II. FDA event 95380 has 2 product records. The first FDA enforcement report date is 30 Oct 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Recalling firm

Olympus Corporation of the Americas, Center Valley, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 95380: product records
Recall numberProductQuantityClassStatusReport date
Z-0134-2025Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, Model/Catalog Number: TFL-PLS1373 unitsClass IIOngoing30 Oct 2024
Z-0135-2025Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS159 unitsClass IIOngoing30 Oct 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
141

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 95380. https://recallregister.com/fda/ii/95380/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
What quantity of product is in FDA event 95380?
The event has 2 product records. The first record gives this quantity: 1373 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.