Recall Register

Device recall: Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR BLUE Model/Catalog Number: 4329BL Product Description…

Class II FDA enforcement event 95375. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Nova Ortho-Med Inc started this device recall on 26 Aug 2024. The firm address in the FDA record is in Carson, CA. FDA put the recall in Class II. FDA event 95375 has 3 product records. The first FDA enforcement report date is 23 Oct 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - U.S. Nationwide distribution including in the states of AK, AL AZ, CA, CO, CT, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK,…

Recalling firm

Nova Ortho-Med Inc, Carson, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 95375: product records
Recall numberProductQuantityClassStatusReport date
Z-0129-2025Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR BLUE Model/Catalog Number: 4329BL Product Description: ROLLING WALKER MOBILITY AID1886Class IIOngoing23 Oct 2024
Z-0130-2025Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR PURPLE Model/Catalog Number: 4329PL Product Description: ROLLING WALKER MOBILITY AID977Class IIOngoing23 Oct 2024
Z-0131-2025Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID1585Class IIOngoing23 Oct 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
3

Other events of the firm

Other FDA enforcement events of Nova Ortho-Med Inc
InitiatedProduct (first record)TypeClassRecords
7 Oct 2019Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)DeviceClass II1
1 Oct 20164205 Journey Rolling Walker Product Usage: used to assist people in walkingDeviceClass II1

Nova Ortho-Med Inc: all 3 events

Official channels

Cite this page

Recall Register. FDA enforcement event 95375. https://recallregister.com/fda/ii/95375/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL AZ, CA, CO, CT, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK,…
What quantity of product is in FDA event 95375?
The event has 3 product records. The first record gives this quantity: 1886
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.