Recall Register

Device recall: The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID)…

Class II FDA enforcement event 95320. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

FUJIFILM Healthcare Americas Corporation started this device recall on 23 Nov 2023. The firm address in the FDA record is in Lexington, MA. FDA put the recall in Class II. FDA event 95320 has 1 product record. The first FDA enforcement report date is 2 Oct 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide and Worldwide

Recalling firm

FUJIFILM Healthcare Americas Corporation, Lexington, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: FDA Mandated.

Products

1 product record in the event.

FDA event 95320: product records
Recall numberProductQuantityClassStatusReport date
Z-3158-2024The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.130Class IIOngoing2 Oct 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
14

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 95320. https://recallregister.com/fda/ii/95320/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide and Worldwide
What quantity of product is in FDA event 95320?
The FDA record gives this quantity: 130
How did the recall start?
The FDA record says: FDA Mandated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.