Device recall: Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is…
Class II FDA enforcement event 95294. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Outset Medical, Inc. started this device recall on 1 Aug 2024. The firm address in the FDA record is in San Jose, CA. FDA put the recall in Class II. FDA event 95294 has 1 product record. The first FDA enforcement report date is 9 Oct 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, MT, OR, ND. OUS: UAE
Recalling firm
Outset Medical, Inc., San Jose, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0019-2025 | Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. | 2,422 | Class II | Ongoing | 9 Oct 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 5
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 30 Apr 2024 | TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 | Device | Class II | 1 |
| 6 Mar 2024 | Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001 | Device | Class I | 1 |
| 24 Jan 2023 | Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system ( | Device | Class II | 1 |
| 20 Jan 2022 | Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0019-2025The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 95294. https://recallregister.com/fda/ii/95294/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, MT, OR, ND. OUS: UAE
- What quantity of product is in FDA event 95294?
- The FDA record gives this quantity:
2,422
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.