Device recall: WALLFLEX FC ESO STENT RMV 18X103MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System…
Class II FDA enforcement event 95124. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
WallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Boston Scientific Corporation started this device recall on 1 Aug 2024. The firm address in the FDA record is in Marlborough, MA. FDA put the recall in Class II. FDA event 95124 has 31 product records. The first FDA enforcement report date is 2 Oct 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide and the countries of APAC, Canada, EMEA, LATAM.
Recalling firm
Boston Scientific Corporation, Marlborough, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
31 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-3265-2024 | WALLFLEX FC ESO STENT RMV 18X103MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 79 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3266-2024 | WALLFLEX FC ESO STENT RMV 18X123MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 66 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3267-2024 | WALLFLEX FC ESO STENT RMV 18X153MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 2 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3268-2024 | WALLFLEX FC ESO STENT RMV 23X105MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 146 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3269-2024 | WALLFLEX FC ESO STENT RMV 23X125MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 176 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3270-2024 | WALLFLEX FC ESO STENT RMV 23X155MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 145 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3271-2024 | WALLFLEX FC ESO STENT RMV LL 18X103- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 77 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3272-2024 | WALLFLEX FC ESO STENT RMV LL 23X105- Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN): M00516300 | 40 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3273-2024 | WALLFLEX FC ESO STENT RMV LL 23X125-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 115 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3274-2024 | WALLFLEX FC ESO STENT RMV LL 23X155- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 77 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3275-2024 | WALLFLEX ESOPHAGEAL FC 18/23-25MMX10CM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number… | 231 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3276-2024 | WALLFLEX ESOPHAGEAL FC 18/23-25MMX12CM-- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent System Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number… | 124 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3277-2024 | WALLFLEX ESOPHAGEAL FC 18/23-25MMX15CM- Agile Esophageal Partially Covered and Fully Covered OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 79 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3278-2024 | WALLFLEX ESOPHAGEAL FC 23/28MMX10CM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 364 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3279-2024 | WALLFLEX ESOPHAGEAL FC 23/28MMX12CM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 507 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3280-2024 | WALLFLEX ESOPHAGEAL FC 23/28MMX15CM-Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN): M00516750 | 516 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3281-2024 | WALLFLEX PC ESOPH STENT 18/23MM X 103MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 297 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3282-2024 | WALLFLEX PC ESOPH STENT 18/23MM X 123MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 308 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3283-2024 | WALLFLEX PC ESOPH STENT 18/23MM X 153MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 39 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3284-2024 | WALLFLEX PC ESOPH STENT 23/28MM X 105MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 133 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3285-2024 | WALLFLEX PC ESOPH STENT 23/28MM X 125MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 305 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3286-2024 | WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas Material Number (UPN):… | 199 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3287-2024 | AGILE ESO OTW PC 23MM X 6.2CM IN 18.5 FR- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 6 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3288-2024 | AGILE ESO OTW PC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 15 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3289-2024 | AGILE ESO OTW PC 23MM X 12.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 17 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3290-2024 | AGILE ESO OTW PC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 18 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3291-2024 | AGILE ESO OTW FC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 18 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3292-2024 | AGILE ESO OTW FC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 19 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3293-2024 | AGILE ESO FC RMV 23MM X 6.2CM IN 18.5 FR-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 7 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3294-2024 | AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 10 units | Class II | Ongoing | 2 Oct 2024 |
| Z-3295-2024 | AGILE ESO FC RMV 23MM X 15.0CM IN 18.5 F- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. Material… | 6 units | Class II | Ongoing | 2 Oct 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 31Median of the firm: 2
- Events of the firm on the register
- 162
Other events of the firm
Official channels
- FDA enforcement report Z-3265-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 95124. https://recallregister.com/fda/ii/95124/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide and the countries of APAC, Canada, EMEA, LATAM.
- What quantity of product is in FDA event 95124?
- The event has 31 product records. The first record gives this quantity:
79 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.