Recall Register

Device recall: Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The…

Class II FDA enforcement event 94928. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Hitachi America, Ltd., Particle Therapy Division started this device recall on 27 May 2024. The firm address in the FDA record is in Santa Clara, CA. FDA put the recall in Class II. FDA event 94928 has 3 product records. The first FDA enforcement report date is 18 Sep 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S. Nationwide distribution in the states of TN, TX, and WA.

Recalling firm

Hitachi America, Ltd., Particle Therapy Division, Santa Clara, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 94928: product records
Recall numberProductQuantityClassStatusReport date
Z-3125-2024Brand Name: PROBEAT Product Name: PROBEAT-FR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and…1Class IIOngoing18 Sep 2024
Z-3126-2024Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and…1Class IIOngoing18 Sep 2024
Z-3127-2024Brand Name: PROBEAT Product Name: PROBEAT-V Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and…1Class IIOngoing18 Sep 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
3
Events of the firm on the register
1

Official channels

Cite this page

Recall Register. FDA enforcement event 94928. https://recallregister.com/fda/ii/94928/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S. Nationwide distribution in the states of TN, TX, and WA.
What quantity of product is in FDA event 94928?
The event has 3 product records. The first record gives this quantity: 1
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.