Device recall: ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and…

Class II FDA enforcement event 94892. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens AG/Siemens Healthcare GmbH started this device recall on 26 Jun 2024. The firm address in the FDA record is in Forchheim. FDA put the recall in Class II. FDA event 94892 has 3 product records. The first FDA enforcement report date is 10 Jul 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.

Recalling firm

Siemens AG/Siemens Healthcare GmbH, Forchheim, N/A, Germany.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: FDA Mandated.

Products

3 product records in the event.

FDA event 94892: product records
Recall numberProductQuantityClassStatusReport date
Z-2240-2024ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.201 units in totalClass IIOngoing10 Jul 2024
Z-2241-2024ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.201 units in totalClass IIOngoing10 Jul 2024
Z-2242-2024ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.201 units in totalClass IIOngoing10 Jul 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
3
Events of the firm on the register
1

Official channels

Cite this page

Recall Register. FDA enforcement event 94892. https://recallregister.com/fda/ii/94892/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
What quantity of product is in FDA event 94892?
The event has 3 product records. The first record gives this quantity: 201 units in total
How did the recall start?
The FDA record says: FDA Mandated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.