Recall Register

Device recall: Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product…

Class II FDA enforcement event 94890. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

B Braun Medical Inc started this device recall on 14 Jun 2024. The firm address in the FDA record is in Breinigsville, PA. FDA put the recall in Class II. FDA event 94890 has 5 product records. The first FDA enforcement report date is 31 Jul 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Domestic distribution nationwide. International distribution to Canada.

Recalling firm

B Braun Medical Inc, Breinigsville, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 94890: product records
Recall numberProductQuantityClassStatusReport date
Z-2433-2024Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A14949Class IIOngoing31 Jul 2024
Z-2434-2024Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A9943Class IIOngoing31 Jul 2024
Z-2435-2024Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A1663Class IIOngoing31 Jul 2024
Z-2436-2024Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A930Class IIOngoing31 Jul 2024
Z-2437-2024Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A4519Class IIOngoing31 Jul 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
5Median of the firm: 1
Events of the firm on the register
90

Other events of the firm

Other FDA enforcement events of B Braun Medical Inc
InitiatedProduct (first record)TypeClassRecords
19 Aug 20260.9% Sodium Chloride Injection, USP, 100 mL Partial Fill in 150 mL PAB Container, Sterile, Rx only, B. Braun Medical IncDrugClass I1
12 Aug 2026Lactated Ringer's Injection USP, 1000 mL EXCEL container, Rx only, L7500, B. Braun Medical, Inc., Bethlehem, PA 18018-35DrugClass I1
4 Aug 2026Brand Name: Extension Set Product Name: DISCOFIX FOUR WAY STOPCOCK Model/Catalog Number: 456025DeviceClass II1
6 May 2026Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); DeviceClass I3
28 Apr 2026Lactated Ringer's Injection USP, 1000 mL container, Rx only, B. Braun Medical, Inc., Bethlehem, PA 18018-3524 USA, NDC 0DrugClass I1
2 Apr 2026Sterile Water for Injection USP, 3000 mL bags, Rx only, Pharmacy Bulk Package, B. Braun Medical Inc., Bethlehem, PA 1801DrugClass II4
19 Mar 2026B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.DeviceClass I6
2 Feb 2026Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version: NDeviceClass II3
29 Jan 2026Brand Name: B. BRAUN MEDICAL INC Product Name: CESK NORTHSIDE ANESTHESIA TRAY Model/Catalog Number: 555076 Software VersDeviceClass II8
29 Oct 2025ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space LarDeviceClass II116

B Braun Medical Inc: all 90 events

Official channels

Cite this page

Recall Register. FDA enforcement event 94890. https://recallregister.com/fda/ii/94890/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Domestic distribution nationwide. International distribution to Canada.
What quantity of product is in FDA event 94890?
The event has 5 product records. The first record gives this quantity: 14949
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.