Device recall: DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Class II FDA enforcement event 94840. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Beckman Coulter, Inc. started this device recall on 22 May 2024. The firm address in the FDA record is in Chaska, MN. FDA put the recall in Class II. FDA event 94840 has 1 product record. The first FDA enforcement report date is 4 Sep 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt,…
Recalling firm
Beckman Coulter, Inc., Chaska, MN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2735-2024 | DxI 9000 Access Immunoassay Analyzer, Catalog #C11137. | 183 analyzers | Class II | Ongoing | 4 Sep 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 216
Other events of the firm
Official channels
- FDA enforcement report Z-2735-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94840. https://recallregister.com/fda/ii/94840/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, TX, and WA. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czechia, Egypt,…
- What quantity of product is in FDA event 94840?
- The FDA record gives this quantity:
183 analyzers
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.