Recall Register

Device recall: ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating…

Class II FDA enforcement event 94792. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Zoll Medical Corporation started this device recall on 31 May 2024. The firm address in the FDA record is in Deerfield, WI. FDA put the recall in Class II. FDA event 94792 has 1 product record. The first FDA enforcement report date is 17 Jul 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR

Recalling firm

Zoll Medical Corporation, Deerfield, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 94792: product records
Recall numberProductQuantityClassStatusReport date
Z-2324-2024ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C…14885Class IIOngoing17 Jul 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
8

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 94792. https://recallregister.com/fda/ii/94792/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR
What quantity of product is in FDA event 94792?
The FDA record gives this quantity: 14885
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.