Device recall: VidiaPort carrying arm, mount for auxiliary products, model numbers: a) 4028110 (Ceiling Single Surgical Light); b)…
Class II FDA enforcement event 94766. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Baxter Healthcare Corporation started this device recall on 4 Jun 2024. The firm address in the FDA record is in Deerfield, IL. FDA put the recall in Class II. FDA event 94766 has 3 product records. The first FDA enforcement report date is 17 Jul 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution.
Recalling firm
Baxter Healthcare Corporation, Deerfield, IL.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2313-2024 | VidiaPort carrying arm, mount for auxiliary products, model numbers: a) 4028110 (Ceiling Single Surgical Light); b) 4028210 (Ceiling Duo Surgical Lights); c) 4028310 (Ceiling Trio Surgical Lights) | 275 units | Class II | Ongoing | 17 Jul 2024 |
| Z-2314-2024 | Trulight 5000/3000 Surgical Light, model numbers: a) 4038210 (Duo Surgical Lights}; b) 4038310 (Trio/Quad Surgical Lights) | 5 units | Class II | Ongoing | 17 Jul 2024 |
| Z-2315-2024 | iLED 7 surgical lights, Model Numbers: a) 4068110 (Ceiling Single Surgical Light); b) 4068120 (Mobile Surgical Light); c) 4068140 (Pendant Surgical Light); d) 4068210 (Ceiling Duo Surgical Lights); e) 4068310 (Ceiling Trio Surgical Lights); f) 4068410 (Ceiling Quad Surgical Lights) | 6243 units | Class II | Ongoing | 17 Jul 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 193
Other events of the firm
Official channels
- FDA enforcement report Z-2313-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94766. https://recallregister.com/fda/ii/94766/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution.
- What quantity of product is in FDA event 94766?
- The event has 3 product records. The first record gives this quantity:
275 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?
- 87% of the 970 device events that started in 2024 are in Class II.