Recall Register

Device recall: 20 Gauge V-Lance Knife Model/Catalog Number: 8065912001 Software Version: N/A Product Description: Ophthalmic knife…

Class II FDA enforcement event 94501. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Alcon Research LLC started this device recall on 22 Apr 2024. The firm address in the FDA record is in Fort Worth, TX. FDA put the recall in Class II. FDA event 94501 has 4 product records. The first FDA enforcement report date is 5 Jun 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany,…

Recalling firm

Alcon Research LLC, Fort Worth, TX.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 94501: product records
Recall numberProductQuantityClassStatusReport date
Z-1922-202420 Gauge V-Lance Knife Model/Catalog Number: 8065912001 Software Version: N/A Product Description: Ophthalmic knife Component: N/A29,190 unitsClass IIOngoing5 Jun 2024
Z-1923-2024ClearCut Sideport Knife Dual Bevel 1.2mm Angled Model/Catalog Number: 8065921541 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A427,764 unitsClass IIOngoing5 Jun 2024
Z-1924-2024ClearCut S Safety Sideport Knife 1.0mm Dual Bevel Model/Catalog Number: 8065771540 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A19,867 unitsClass IIOngoing5 Jun 2024
Z-1925-2024ClearCut S Safety Sideport Knife 1.2mm Dual Bevel Model/Catalog Number: 8065771541 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A4,930 unitsClass IIOngoing5 Jun 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
4Median of the firm: 1
Events of the firm on the register
25

Other events of the firm

Other FDA enforcement events of Alcon Research LLC
InitiatedProduct (first record)TypeClassRecords
21 Apr 2026Systane, Lubricant Eye Gel, Night Gel, Sterile, 10g (0.35 oz), Processed in France for: Alcon Laboratiroes, Inc., Fort WDrugClass II2
24 Nov 2025Brand Name: Alcon Laboratories, Inc. Product Name: Custom Pak Model/Catalog Numbers: 9903-20, 17188-04, 19762-03, 19203-DeviceClass I1
30 Sep 2025Alcon PRECISION1 One-Day Contact Lenses, -2.25D.DeviceClass II1
15 Sep 2025CONSTELLATION ULTRAVIT 10K and CONSTELLATION HYPERVIT 20K, Description/REF: 25+ TOTALPLUS CP PAK 20K CPM BWV .9 IU/ 8065DeviceClass II1
5 Aug 2025Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles ModelDeviceClass II5
28 Apr 2025Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Model/Catalog NuDeviceClass II3
10 Feb 2025Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-DeviceClass II1
18 Dec 2024Systane Lubricant Eye Drops, Ultra PF, Sterile, 25 Vials (0.7mL Each), Manufactured for: Alcon Laboratories, Inc. Forth DrugClass I1
13 Nov 2024Brand Name: Alcon Product Name: Monarch III D Delivery Cartridge Model/Catalog Number: 8065977763 Product Description: MDeviceClass II1
15 Oct 2024a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALEM SUMP, 14 FR; DeviceClass III1

Alcon Research LLC: all 25 events

Official channels

Cite this page

Recall Register. FDA enforcement event 94501. https://recallregister.com/fda/ii/94501/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany,…
What quantity of product is in FDA event 94501?
The event has 4 product records. The first record gives this quantity: 29,190 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.