Device recall: PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S…
Class II FDA enforcement event 94257. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
SPINEART SA started this device recall on 15 Mar 2024. The firm address in the FDA record is in Plan-Les-Ouates. FDA put the recall in Class II. FDA event 94257 has 1 product record. The first FDA enforcement report date is 15 May 2024. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
Recalling firm
SPINEART SA, Plan-Les-Ouates, N/A, Switzerland.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1758-2024 | PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA TL system is intended to provide immobilization and stabilization of spinal segments in… | 242 systems | Class II | Ongoing | 15 May 2024 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2024
- 970
- Share of these events in Class II
- 87%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 6
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 30 Jun 2025 | Brand Name: PERLA TL MIS Product Name: CANNULATED FENESTRATED POLYAXIAL SCREW Model/Catalog Number: MPF-PS 65 40-S Softw | Device | Class II | 2 |
| 16 Mar 2023 | JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JLT-PL 02 14-S, J | Device | Class II | 1 |
| 9 Apr 2021 | PERLA TL Lateral Connector Open - Product Usage: intended to be used with autograft and/or allograft. Pediatric pedicle | Device | Class II | 1 |
| 5 Feb 2019 | Intersomatic JULIET Ti OL cage. For use in orthopedic spinal procedures | Device | Class III | 1 |
| 27 Nov 2017 | TRYPTIK(R) 2 C-PLATE SELF-TAPPING POLYAXIAL SCREW 4.0 MM, LENGTH 12 MM, REF TRY-PS-40 12-S, Qty 2 TRYPTIK(R) 2C-Plate An | Device | Class II | 7 |
Official channels
- FDA enforcement report Z-1758-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 94257. https://recallregister.com/fda/ii/94257/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
- What quantity of product is in FDA event 94257?
- The FDA record gives this quantity:
242 systems
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2024 have the same class?