Recall Register

Device recall: EndoVive 20Fr Push Safety PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both…

Class II FDA enforcement event 94165. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Boston Scientific Corporation started this device recall on 20 Feb 2024. The firm address in the FDA record is in Marlborough, MA. FDA put the recall in Class II. FDA event 94165 has 5 product records. The first FDA enforcement report date is 17 Apr 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the country of Japan.

Recalling firm

Boston Scientific Corporation, Marlborough, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 94165: product records
Recall numberProductQuantityClassStatusReport date
Z-1513-2024EndoVive 20Fr Push Safety PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M005090112726 unitsClass IIOngoing17 Apr 2024
Z-1514-2024EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00509051572 unitsClass IIOngoing17 Apr 2024
Z-1515-2024EndoVive 20Fr Push Safety PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M005664712501 unitsClass IIOngoing17 Apr 2024
Z-1516-2024EndoVive 20Fr Push Standard PEG Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M005682111475 unitsClass IIOngoing17 Apr 2024
Z-1517-2024EndoVive 20 Fr Safety PEG Push w/ENFit Kit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M0050919113 units (Japan)Class IIOngoing17 Apr 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
5Median of the firm: 2
Events of the firm on the register
162

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 94165. https://recallregister.com/fda/ii/94165/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the country of Japan.
What quantity of product is in FDA event 94165?
The event has 5 product records. The first record gives this quantity: 2726 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.