Recall Register

Device recall: InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinus surgery.…

Class II FDA enforcement event 94067. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Japanese IFU version had slightly different language and illustrations from the English IFU, which could contribute to the tip and plate falling off. In order to maintain consistency globally, the Japanese IFU will be updated to reflect the instructions in the English IFU

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Olympus Corporation of the Americas started this device recall on 16 Jan 2024. The firm address in the FDA record is in Center Valley, PA. FDA put the recall in Class II. FDA event 94067 has 1 product record. The first FDA enforcement report date is 3 Apr 2024. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

International distribution to the country of Japan.

Recalling firm

Olympus Corporation of the Americas, Center Valley, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 94067: product records
Recall numberProductQuantityClassStatusReport date
Z-1368-2024InstaClear Lens Cleaner-indicated for use during routine diagnostic procedures and during endoscopic sinus surgery. Model Number: LCS1500UNOL LCS1800UNST LCS4K00UNOL LCS1530BTOL LCS1830BTST LCS4K30BTOL LCS1530TPOL LCS1830TPST LCS4K30TPOL LCS1545BTOL LCS1845BTST LCS4K45BTOL LCS1545TPOL LCS1845TPST LCS4K45TPOL…102,279 units (OUS Only)Class IIOngoing3 Apr 2024

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2024
970
Share of these events in Class II
87%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
141

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 94067. https://recallregister.com/fda/ii/94067/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: International distribution to the country of Japan.
What quantity of product is in FDA event 94067?
The FDA record gives this quantity: 102,279 units (OUS Only)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2024 have the same class?
87% of the 970 device events that started in 2024 are in Class II.