Recall Register

Device recall: Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780…

Class II FDA enforcement event 93563. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Maquet Medical Systems USA started this device recall on 13 Nov 2023. The firm address in the FDA record is in Wayne, NJ. FDA put the recall in Class II. FDA event 93563 has 1 product record. The first FDA enforcement report date is 27 Dec 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark,…

Recalling firm

Maquet Medical Systems USA, Wayne, NJ.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 93563: product records
Recall numberProductQuantityClassStatusReport date
Z-0605-2024Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 7010418021213 US, 2589 OUSClass IIOngoing27 Dec 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
20

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 93563. https://recallregister.com/fda/ii/93563/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark,…
What quantity of product is in FDA event 93563?
The FDA record gives this quantity: 1213 US, 2589 OUS
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.