Recall Register

Device recall: Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.

Class II FDA enforcement event 93406. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Pump not detected as connected to controller due to software issue.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Abiomed, Inc. started this device recall on 10 Oct 2023. The firm address in the FDA record is in Danvers, MA. FDA put the recall in Class II. FDA event 93406 has 2 product records. The first FDA enforcement report date is 13 Dec 2023. The record gives the status Terminated, with the termination date 6 Mar 2025.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US: AZ, CA, FL, NY & VA.

Recalling firm

Abiomed, Inc., Danvers, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 93406: product records
Recall numberProductQuantityClassStatusReport date
Z-0440-2024Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.Class IITerminated13 Dec 2023
Z-0441-2024Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.Class IITerminated13 Dec 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
30

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 93406. https://recallregister.com/fda/ii/93406/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US: AZ, CA, FL, NY & VA.
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.
How many FDA recalls does Abiomed, Inc. have?
Abiomed, Inc. is the recalling firm in 30 enforcement events on this register. This event is 1 of them.