Device recall: ELEKTA LIMITED Precise Digital Accelerator, REF MRT 6001, MRT 6011; Medical Charged-Particle Radiation Therapy System…
Class II FDA enforcement event 93264. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Elekta, Inc. started this device recall on 25 Oct 2023. The firm address in the FDA record is in Atlanta, GA. FDA put the recall in Class II. FDA event 93264 has 6 product records. The first FDA enforcement report date is 13 Dec 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
worldwide
Recalling firm
Elekta, Inc., Atlanta, GA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
6 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0442-2024 | ELEKTA LIMITED Precise Digital Accelerator, REF MRT 6001, MRT 6011; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 315 units | Class II | Ongoing | 13 Dec 2023 |
| Z-0443-2024 | ELEKTA LIMITED Elekta Synergy, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 1019 units | Class II | Ongoing | 13 Dec 2023 |
| Z-0444-2024 | ELEKTA LIMITED Elekta Synergy Platform, REF MRT 8501, MRT 8511; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 264 units | Class II | Ongoing | 13 Dec 2023 |
| Z-0445-2024 | ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 451 units | Class II | Ongoing | 13 Dec 2023 |
| Z-0446-2024 | ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 291 units | Class II | Ongoing | 13 Dec 2023 |
| Z-0447-2024 | ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear | 47 units | Class II | Ongoing | 13 Dec 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 6Median of the firm: 1
- Events of the firm on the register
- 105
Other events of the firm
Official channels
- FDA enforcement report Z-0442-2024The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 93264. https://recallregister.com/fda/ii/93264/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
worldwide
- What quantity of product is in FDA event 93264?
- The event has 6 product records. The first record gives this quantity:
315 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?