Device recall: Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN…
Class II FDA enforcement event 92893. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Olympus Corporation of the Americas started this device recall on 10 Aug 2023. The firm address in the FDA record is in Center Valley, PA. FDA put the recall in Class II. FDA event 92893 has 6 product records. The first FDA enforcement report date is 11 Oct 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
6 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2659-2023 | Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 0095 SYS-3000 SPiN Thoracic Navigation System SYS-2400 SPiN Drive System SYS-2230 ig4 System V2, 230V, 50Hz… | 226 units | Class II | Ongoing | 11 Oct 2023 |
| Z-2660-2023 | Veran Endobronchial: Models: INS-5925 SPiN Access Catheter 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter… | 6484 units | Class II | Ongoing | 11 Oct 2023 |
| Z-2661-2023 | Veran : Percutaneous Models: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" INS-5611 SPiN Perc Localization Needle - 2cm, 5/Bx INS-5610 SPiN Perc Localization Needle - 1cm, 5/Bx INS-5600 SPiN Perc Biopsy Needle Guide Kit, 5/Bx (includes INS-5029) | 9961 units | Class II | Ongoing | 11 Oct 2023 |
| Z-2662-2023 | Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC INS-5411 0.016" Stylet, 1235 mm L, 10/Bx INS-5410 SPiN Flex ANSO Needle, 22ga, 1.8mm OD, 5/Bx INS-5323 Sliding stopper INS-5300 "Always-On Tip Tracked Triple… | 6318 units | Class II | Ongoing | 11 Oct 2023 |
| Z-2663-2023 | Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point… | 522 units | Class II | Ongoing | 11 Oct 2023 |
| Z-2664-2023 | Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322… | 60102 units | Class II | Ongoing | 11 Oct 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 6Median of the firm: 1
- Events of the firm on the register
- 141
Other events of the firm
Official channels
- FDA enforcement report Z-2659-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 92893. https://recallregister.com/fda/ii/92893/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
- What quantity of product is in FDA event 92893?
- The event has 6 product records. The first record gives this quantity:
226 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?
- 88% of the 917 device events that started in 2023 are in Class II.