Recall Register

Device recall: API3200MD Mass Spectrometer, Part Number 4466230

Class II FDA enforcement event 92810. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

AB Sciex, LLC started this device recall on 30 Aug 2023. The firm address in the FDA record is in Framingham, MA. FDA put the recall in Class II. FDA event 92810 has 4 product records. The first FDA enforcement report date is 11 Oct 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide.

Recalling firm

AB Sciex, LLC, Framingham, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

4 product records in the event.

FDA event 92810: product records
Recall numberProductQuantityClassStatusReport date
Z-0037-2024API3200MD Mass Spectrometer, Part Number 44662303Class IIOngoing11 Oct 2023
Z-0038-20244500MD Triple Quad Mass Spectrometer, Part Number 503252238Class IIOngoing11 Oct 2023
Z-0039-2024Citrine QTRAP Mass Spectrometer, Part Number 506368510Class IIOngoing11 Oct 2023
Z-0040-2024Citrine Triple Quad Mass Spectrometer, Part Number 506368414Class IIOngoing11 Oct 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
4
Events of the firm on the register
1

Official channels

Cite this page

Recall Register. FDA enforcement event 92810. https://recallregister.com/fda/ii/92810/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide.
What quantity of product is in FDA event 92810?
The event has 4 product records. The first record gives this quantity: 3
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.