Recall Register

Device recall: Flexiva Pulse 242 Single-Use Laser Fibers-intended to be used as a device that transmits Ho:YAG laser energy from…

Class II FDA enforcement event 92570. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Boston Scientific Corporation started this device recall on 25 May 2023. The firm address in the FDA record is in Marlborough, MA. FDA put the recall in Class II. FDA event 92570 has 7 product records. The first FDA enforcement report date is 2 Aug 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the country of Canada.

Recalling firm

Boston Scientific Corporation, Marlborough, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

7 product records in the event.

FDA event 92570: product records
Recall numberProductQuantityClassStatusReport date
Z-2228-2023Flexiva Pulse 242 Single-Use Laser Fibers-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. Single-pack UPN : M006L840591066 unitsClass IIOngoing2 Aug 2023
Z-2229-2023Flexiva Pulse 242 TracTip Single-Use Laser Fibers - intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Single-pack UPN : M006L8405960339 unitsClass IIOngoing2 Aug 2023
Z-2230-2023Flexiva Pulse 242 TracTip Single-Use Laser Fibers -intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. 5-pack UPN: M006L84059611245 unitsClass IIOngoing2 Aug 2023
Z-2231-2023Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Single-pack UPN: M006L8406910112 unitsClass IIOngoing2 Aug 2023
Z-2232-2023Flexiva Pulse ID-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Box of 5)UPN: M006L8406911205 unitsClass IIOngoing2 Aug 2023
Z-2233-2023Flexiva Pulse 242 ID Trac Single Use Fiber-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy (Single-pack UPN): M006L8406960231 unitsClass IIOngoing2 Aug 2023
Z-2234-2023Flexiva Pulse ID 242 TracTip single Use Fiber-intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy 5-pack UPN: M006L8406961280 unitsClass IIOngoing2 Aug 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
7Median of the firm: 2
Events of the firm on the register
162

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 92570. https://recallregister.com/fda/ii/92570/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the country of Canada.
What quantity of product is in FDA event 92570?
The event has 7 product records. The first record gives this quantity: 66 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.