Device recall: Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30 Blade…
Class II FDA enforcement event 92297. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Packaging for a serrated blade applicator kit (12 Sample) may contain a serrated blade applicator (18 sample). This results in a characteristic pattern of missed or light samples on the gel, leading to incorrect or no result, or delay in result.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Helena Laboratories, Corp. started this device recall on 2 May 2023. The firm address in the FDA record is in Beaumont, TX. FDA put the recall in Class II. FDA event 92297 has 1 product record. The first FDA enforcement report date is 28 Jun 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Distribution to US states of PA, NE, MI, FL, AR, NY, TX, and CA, and to Thailand.
Recalling firm
Helena Laboratories, Corp., Beaumont, TX.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1975-2023 | Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30 Blade Applicators. used in invitro diagnostics | 32 units | Class II | Ongoing | 28 Jun 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 9
Other events of the firm
Official channels
- FDA enforcement report Z-1975-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 92297. https://recallregister.com/fda/ii/92297/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Distribution to US states of PA, NE, MI, FL, AR, NY, TX, and CA, and to Thailand.
- What quantity of product is in FDA event 92297?
- The FDA record gives this quantity:
32 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?